What do I do if the bead in the flow meter is sticking/not moving?

What do I do if the bead in the flow meter is sticking/not moving?

If the Pumani unit is not in use for some time, or is used in a very humid environment, the bead inside of the flow meter can become stuck.

To loosen the bead, you will need to take the following steps:

1.       Open the unit

2.       Identify the faulty flow meter

3.       Remove the tubing from the inlet and outlet of the flow meter by pulling forcefully on the tubing until it removes from the barbed connector (do NOT cut the tubing)

4.       Unscrew the two screws holding the faulty flow meter to the enclosure

5.       Separate the flow meter from the enclosure

6.       Turn the flow meter upside down and shake it until you hear that the bead is loose and rattles in the meter

7.       Place the meter back into place on the enclosure

8.       Reattach the two screws to the flow meter

9.       Reattach the tubing to the flow meter, making sure the tubing is still attached correctly

10.   Close the unit

11.   Turn on the unit and confirm that the bead is no longer stuck in the flow meter


    • Related Articles

    • What do I do if the oxygen flow meter is not working?

      If the bead in the oxygen flow meter is not moving, follow the troubleshooting steps below: Turn on the unit Ensure the oxygen flow meter is fully open Attach external oxygen source  Gently knock on the oxygen flow meter If the bead is still not ...
    • What do I do if the knob on one of the flow meter detaches?

      If you loosen a flow meter too much, the knob may detach from the meter. If the knob detaches, you can place the knob back on the meter by turning the knob to the right until it reattaches into the meter.
    • What do I do if the total flow is too low?

      If the flow is too low, you may need to replace the filter under the Pumani pump. For instructions on how to check the flow and how to replace the filter, see the User & Repair Manual.
    • What should I do if I drop the unit?

      The Pumani is designed to withstand drops from typical heights. If you do drop the unit, place the unit back on a level surface, check for any external damage to the unit, plug in and turn on the unit, and make sure the indicator light turns on, the ...
    • What do I do if the oxygen tubing detaches from the oxygen port?

      The oxygen tubing may detach from the oxygen port if the oxygen flow is much higher than the total flow, or if the flow meters are closed and the oxygen flow cannot be released. To prevent this from happening, make sure the flow meters are open and ...